您的位置: 专家智库 > >

国家自然科学基金(s30700217)

作品数:1 被引量:0H指数:0
发文基金:国家自然科学基金国家重点基础研究发展计划更多>>
相关领域:医药卫生更多>>

文献类型

  • 1篇中文期刊文章

领域

  • 1篇医药卫生

主题

  • 1篇OPTIC_...
  • 1篇PENETR...
  • 1篇PLATIN...
  • 1篇VIVO
  • 1篇MICROE...
  • 1篇IRIDIU...

传媒

  • 1篇Journa...

年份

  • 1篇2011
1 条 记 录,以下是 1-1
排序方式:
In-Vitro and In-Vivo Electrical Characteristics of a Penetrating Microelectrode Array for Optic Nerve Electrical Stimulation
2011年
The Chinese C-Sight team aims to restore vision to blind patients by means of stimulating the optic nerve with a penetrating microelectrode array. A biocompatible, implantable microwire array was developed having four platinum-iridium shafts, each 100μm in diameter. This penetrating microwire array is described in this paper, including its fabrication techniques and its in-vitro electrical characteristics. Every set of four shafts was spaced 0.4mm from center to center, comprising two short shafts that were 0.3mm long and two that were 0.9mm long. This design was intended to stimulate ganglion cell axons at different depths within the optic nerve. In-vitro electrochemical impedance testing results showed that the impedance at 1kHz ranged from 8 to 10kΩ at room temperature. The voltage responses of the arrays to current pulse stimulation indicated a charge-injection capacity of 210μC/cm2. Finally, in-vivo acute animal experiments showed that the amplitude of the electrically evoked potentials (EEPs) measured in primary visual cortex could be as large as 100 μV upon direct stimulation of the optic nerve.
隋晓红邵轶彬李丽明柴新禹任秋实
共1页<1>
聚类工具0